瀏覽代碼

iio: pressure: mpl115: fix temperature offset sign

According to the datasheet, the resolusion of temperature sensor is
-5.35 counts/C. Temperature ADC is 472 counts at 25C.
(https://www.sparkfun.com/datasheets/Sensors/Pressure/MPL115A1.pdf
NOTE: This is older revision, but this information is removed from the
latest datasheet from nxp somehow)

Temp [C] = (Tadc - 472) / -5.35 + 25
         = (Tadc - 605.750000) * -0.186915888

So the correct offset is -605.750000.

Signed-off-by: Akinobu Mita <akinobu.mita@gmail.com>
Acked-by: Peter Meerwald-Stadler <pmeerw@pmeerw.net>
Cc: <Stable@vger.kernel.org>
Signed-off-by: Jonathan Cameron <jic23@kernel.org>
Akinobu Mita 9 年之前
父節點
當前提交
431386e783
共有 1 個文件被更改,包括 1 次插入1 次删除
  1. 1 1
      drivers/iio/pressure/mpl115.c

+ 1 - 1
drivers/iio/pressure/mpl115.c

@@ -117,7 +117,7 @@ static int mpl115_read_raw(struct iio_dev *indio_dev,
 		*val = ret >> 6;
 		*val = ret >> 6;
 		return IIO_VAL_INT;
 		return IIO_VAL_INT;
 	case IIO_CHAN_INFO_OFFSET:
 	case IIO_CHAN_INFO_OFFSET:
-		*val = 605;
+		*val = -605;
 		*val2 = 750000;
 		*val2 = 750000;
 		return IIO_VAL_INT_PLUS_MICRO;
 		return IIO_VAL_INT_PLUS_MICRO;
 	case IIO_CHAN_INFO_SCALE:
 	case IIO_CHAN_INFO_SCALE: